TU90 — Turbidimeter

Compact inline turbidity sensor using USEPA 180.1 method. Multi-bubble removal ensures precise, continuous measurement with minimal sample volume and economical calibration.

Category: Brand:

Product Overview

TU90 uses a 90° scattered light measurement method applying the US EPA 180.1 method and is equipped with a multi-bubble removal function to precisely measure the turbidity of samples in real time and continuously. The application of a natural flow sample structure, easy-to-maintain sensor and lamp module, safe waterproof structure, and compact size make installation and maintenance easy. In addition, only a very small number of samples are used in real time, and the use of a small amount of solution during calibration is highly economical.

Applications

  • Drinking Water (Tap/Rural)
  • Swimming Pool Water
  • Ultrapure Water
  • Distilled Water
  • Reclaimed Water
  • Industrial Water
  • Power Plants

Specification

Parameter Details
Method USEPA 180.1 Method / 90° Scattered Light Measurement
Measurement Range 0–100 NTU
Sample Temperature 2–60 °C (35.6–143.6 °F)
Flow Rate 0.1–0.8 L/min (0.026–0.211 gal/min)
Flow Rate (Optimal) 0.3 L/min (0.079 gal/min)
Pressure < 0.5 bar
Response Time < 60 sec
Resolution 0.0001 NTU
Accuracy ±1.0%
Repeatability ±2.0%
Calibration 1point (sample) or 2point (zero and sample) / Formazine Standard 180mL
Light Source Tungsten Lamp
Storage Temperature -20–60 °C (-4–140 °F, Original Package), Unlimited
Warranty – Detector 2 years (Semipermanent Use)
Warranty – Lamp 3 years Replacement Recommended
Cleaning < 3 months Recommended in clean water
Calibration Interval < 12 months Recommended in clean water